UK MHRA · UK · 16 Oct 2026 · Guidance

Register medical devices to place on the market

AI-generated summary

All medical devices, including in vitro diagnostic devices and custom-made devices, must be registered with the UK MHRA before placement on the market in Great Britain. Devices must conform to the Medical Devices Regulations 2002. Registration differs for Northern Ireland; since 28 May 2026, most devices there must be registered on the European Database on Medical Devices instead, except custom-made devices which require MHRA registration within 28 days of becoming available. MHRA registration does not constitute approval or endorsement and cannot be claimed in marketing materials.

From UK MHRA

How to register your medical devices with the MHRA for the markets in Great Britain and Northern Ireland.

All medical devices, including in vitro diagnostic devices (IVDs), custom-made devices and systems or procedure packs, must be registered with the MHRA before they can be placed on the market in Great Britain (England, Wales and Scotland).

In Great Britain devices must conform to the Medical Devices Regulations 2002 (SI 2002 No 618, as amended) (UK MDR 2002) as they apply in Great Britain so that they can be placed on the market and registered with the MHRA.

Registration requirements differ for Northern Ireland.

You must ensure all information registered with the MHRA is accurate and up to date.

Registration of your devices with the MHRA (the UK Competent Authority) does not represent any form of accreditation, certification, approval or endorsement by the MHRA.

Therefore, you are not permitted to make any claims to this effect, including the use of any MHRA logos in any marketing materials, on device packaging, in the instructions for use, on laboratory tickets/dockets, or in any other documentation.

As part of our data validation and scrutiny of applications to register medical devices and products with the MHRA, we may request further technical documentation from you to demonstrate that your products conform to the requirements of the Medical Device Regulations.

Registration for the Great Britain Market

The following requirements apply to place your device on the Great Britain market.

It is a requirement of the UK MDR 2002 that you inform the MHRA before you place your device on the market in Great Britain.

You must register if you or your company sells, leases, lends or gifts:

- Class I, IIa, IIb or III devices you have manufactured

- Class I, IIa, IIb or III devices you have refurbished or re-labelled with your own name

- any system or procedure pack containing at least one medical device

- custom-made devices

- IVDs you have manufactured

- IVDs undergoing performance evaluation

If the manufacturer is based outside the UK, they must appoint a single UK Responsible Person to take responsibility for all of their medical devices.

The UK Responsible Person will then assume the responsibilities of the manufacturer in terms of registering the device with the MHRA.

See guidance for more information about the role of the UK Responsible Person.

Where any changes to registrations are made, a statutory fee may apply per application. Please see our Fees section.

Please note that the DORS accounts of any former Great Britain-based Authorised Representatives who had not updated their role to UK Responsible Person by 30 March 2026 are now closed.

Under the UK MDR 2002, “placing on the market” and “putting into service” are defined terms. In most cases, a device will first be placed on the market and will subsequently be put into service when it reaches the end user.

Placing on the market